Summary
Type: SPE
Application #: 305833
Catalog#: 730958 - SPE Cartridges, Chromabond, HR-XA, 6 ml, 150 mg
Detection:
Conditions: Column conditioning: 5 mL methanol, then 5 mL water
(Do not let run the column dry.)
Sample aspiration: prepared sample was applied
Washing: 4 mL water, then 5 mL methanol
Elution: 3 mL 1 mol/L ammonium formate in methanol
Evaporation: under a stream of nitrogen to dryness
Reconstitution: 250 µL water – methanol (1:1, v/v)
Authors: Nagel, R. et al.